| Catalog Number |
305613 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Oem manufacture: the manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified number of sharps coll chemo 19gal yellow experienced missing lids.The following information was provided by the initial reporter: material no: 305613, batch no: 9018920.It was reported that the lids are missing.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: it was reported that lids were missing.No sample photo representation was provided for evaluation.No photo of the original packaging was provided, and the number of missing lids was not confirmed.A review of the device history record (dhr) for the reported lot was performed and there were no issues found for missing lids.A review of non-conforming material report (ncmr) was performed for the last twelve months and did not identify any nonconformance reported for missing lids for the same part number.This lot was confirmed produced after corrective actions.A weight system that provides a weighted label for each final box before shipment has already been implemented for this product to ensure the correct quantity of pieces per box before shipping.It was not possible to confirm that any failure mode was related to the manufacturing process with the information provided since the weight system is validated and did not detect any missing components.There may have potentially been a repackaging issue which occurred during distribution.The actual root cause is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified number of sharps coll chemo 19gal yellow experienced missing lids.The following information was provided by the initial reporter: material no: 305613.Batch no: 9018920.It was reported that the lids are missing.
|
| |
|
Search Alerts/Recalls
|