MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD. 8010379 S-ROM*HEAD FEMORAL COCR 22+0; S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD. 8010379 S-ROM*HEAD FEMORAL COCR 22+0; S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 522022
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); No Code Available (3191)
Date of Event 04/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Hip revision, dislocation.The surgeon also removed from patient, product code 1014-36 but seems the product is not recognized by the system.I¿ll try to attached a document with the original stickers from products.Was surgery delayed due to the reported event? no.Was procedure successfully completed? yes.Patient status/outcome/consequences?no.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #
=
> (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Visual examination of the provided x-ray image confirmed the reported event.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD. 8010379
st. anthony's rd
leeds LS11 8 DT
UK  LS11 8 DT
MDR Report Key8577690
Report Number1818910-2019-92407
Device Sequence Number11147306
Product Code JDI
UDI-Device Identifier10603295170778
UDI-Public10603295170778
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K851422
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number522022
Device Lot NumberS030002
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/12/2019
Supplement Date Received by Manufacturer06/10/2019
Initial Report FDA Received Date05/03/2019
Supplement Report FDA Received Date07/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-