• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code ST Segment Elevation (2059)
Date of Event 03/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a cryo ablation procedure, unexplained st segment elevation was observed on the patient's electrocardiogram (ecg).No treatment was required.Additionally fluid was leaking out of the sheath handle and there was concern the hemostatic valve might be compromised.The sheath was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the sheath 4fc12, with lot number 88765 and the data files were returned and analyzed.The data files indicated eleven applications were performed with balloon catheter 2af284 with lot 21644.Visual inspection of the sheath showed the device was intact with no apparent issues.Air aspiration was reproduced during the pressure test when the dilator was introduced through the sheath.It was noted that the hemostatic valve was leaking, and a torn valve disc was suspected.The reported leaking hemostatic valve was confirmed through analysis.The reported clinical issue (st elevation) could not be confirmed through testing.In conclusion, the sheath failed the return product inspection due to a leaking hemostatic valve.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Common Device Name
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key8578457
Report Number3002648230-2019-00319
Device Sequence Number19249240
Product Code DRA
UDI-Device Identifier00763000043773
UDI-Public00763000043773
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K123591
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/14/2021
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number88765
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2019
Initial Date Received by Manufacturer 04/05/2019
Supplement Date Received by Manufacturer07/17/2019
Initial Report FDA Received Date05/03/2019
Supplement Report FDA Received Date08/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-