| Catalog Number |
4118127090 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Awaiting device evaluation.A follow up report will be filed once the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that the blade broke at the mount during a knee replacement surgery.It was further reported that the piece was retrieved and the procedure was completed with a replacement blade.It was also reported that there were no delays and no adverse consequences as a result of this event.It was further reported that the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results indicate that the bending of the blade coupled with the impact of the blade against metal could have contributed to the breaking of the blade at the mount.The ifu of handpieces associated with this device warn the user against this type of use: "do not apply excessive pressure, such as bending or prying, with the blade.Excessive pressure may bend or fracture the blade and result in tissue damage, loss of tactile control, and/or the ejection of blade fragments at a high velocity".The quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that the blade broke at the mount during a knee replacement surgery.It was further reported that the piece was retrieved and the procedure was completed with a replacement blade.It was also reported that there were no delays and no adverse consequences as a result of this event.It was further reported that the procedure was completed successfully.
|
| |
|