MASIMO - 40 PARKER RD SET NEOPT-500; OXIMETER Back to Search Results
Model Number 4005-9
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event or Problem Description
The customer reported inaccurate low spo2 reading.No patient impact or consequences were reported.
 
Additional Manufacturer Narrative
Additional manufacuring narrative: the returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing.  during simulation testing, the sensor passed manual and preset conditions and provided accurate measurements.The sensor was determined to be functioning as designed.  , other, other text: , corrected data: d4 model # was corrected from 4005 to 4005-9.D4 lot no was corrected from e18gbx to e16kpk.D4 expiration date was corrected from 07/01/2021 to 10/01/2019.H4 device manufacturer date was corrected from 07/02/2018 to 10/07/2016.
 
Event or Problem Description
The customer reported inaccurate low spo2 reading.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RD SET NEOPT-500
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key8580168
Report Number2031172-2019-00233
Device Sequence Number1836563
Product Code DQA
UDI-Device Identifier10843997009895
UDI-Public10843997009895
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K042536
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2019
Device Model Number4005-9
Device Catalogue Number4005
Device Lot NumberE16KPK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/11/2019
Supplement Date Received by Manufacturer06/19/2019
Initial Report FDA Received Date05/03/2019
Supplement Report FDA Received Date07/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-