| Model Number |
4005-9 |
| Medical Device Problem Code |
Low Readings (2460)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported inaccurate low spo2 reading.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacuring narrative: the returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing. during simulation testing, the sensor passed manual and preset conditions and provided accurate measurements.The sensor was determined to be functioning as designed. , other, other text: , corrected data: d4 model # was corrected from 4005 to 4005-9.D4 lot no was corrected from e18gbx to e16kpk.D4 expiration date was corrected from 07/01/2021 to 10/01/2019.H4 device manufacturer date was corrected from 07/02/2018 to 10/07/2016.
|
| |
|
Event or Problem Description
|
|
The customer reported inaccurate low spo2 reading.No patient impact or consequences were reported.
|
| |
|