Model Number 9004
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Reocclusion (1985)
Date of Event 02/22/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis: the sample was not returned from the user facility; therefore, a device evaluation is unable to be performed.A lot history review revealed this is the only complaint associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual sample was not returned for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was not related to the study device or procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.
 
Event or Problem Description
It was reported through a clinical registry that during the index procedure, a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter was used to treat the target lesion located in right proximal popliteal artery.Approximately 10 months after the index procedure, the patient¿s right proximal popliteal artery was reportedly reoccluded via duplex ultrasound (dus) imaging.Approximately 28 months after the index procedure, a revascularization was performed and the health care professional (hcp) deemed it was successful.The investigator assessed the event was not related to the study device or to the procedure.The sample was discarded by the user facility and is not available for further evaluation.No adverse patient effects were reported.
 
Event or Problem Description
It was reported through a clinical registry that during the index procedure, a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter was used to treat the target lesion located in right proximal popliteal artery.Approximately 10 months after the index procedure, the patient¿s right proximal popliteal artery was reportedly reoccluded via duplex ultrasound (dus) imaging.Approximately 28 months after the index procedure, a revascularization was performed and the health care professional (hcp) deemed it was successful.The investigator assessed the event was not related to the study device or to the procedure.The sample was discarded by the user facility and is not available for further evaluation.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
Corrected data: relevant tests/laboratory data.The relevant tests and lab data had the non-target vessel revascularization date changed from "on (b)(6) 2019" to "on (b)(6) 2018".Device evaluated by mfr analysis: the sample was not returned from the user facility; therefore, a device evaluation is unable to be performed.A lot history review revealed this is the only complaint associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual sample was not returned for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was not related to the study device or procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035 DRUG COATED BALLOON PTA CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
MDR Report Key8584350
Report Number3006513822-2019-00073
Device Sequence Number11266781
Product Code ONU
UDI-Device Identifier00801741088605
UDI-Public(01)00801741088605
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,stu
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/01/2017
Device Model Number9004
Device Catalogue NumberLX35130660
Device Lot NumberGFZJ0195
Was Device Available for Evaluation? No
Event Location Hospital
Initial Date Received by Manufacturer 04/11/2019
Supplement Date Received by Manufacturer06/03/2019
Initial Report FDA Received Date05/06/2019
Supplement Report FDA Received Date06/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient Weight97
-
-