BAXTER HEALTHCARE ¿ ROUND LAKE 1000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938739
Device Problem Component Missing (2306)
Patient Problem No Patient Involvement (2645)
Event Date 04/02/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that three (3) 1000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bags were missing the administration port.This was noted prior to product use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was manufactured between july 06, 2018 - july 08, 2018.The actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed to the photograph using the naked eye which revealed the spike port cap was detached.The reported condition was verified.By the nature of the sample, no additional tests were performed.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE ¿ ROUND LAKE
round lake IL
Manufacturer (Section G)
BAXTER HEALTHCARE ¿ ROUND LAKE
25212 w. illinois route 120
round lake IL 60073
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
MDR Report Key8584617
MDR Text Key144853571
Report Number1416980-2019-02559
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477237
UDI-Public(01)00085412477237
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2021
Device Catalogue NumberH938739
Device Lot Number60135986
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/11/2019
Initial Date FDA Received05/06/2019
Supplement Dates Manufacturer Received05/29/2019
Supplement Dates FDA Received06/07/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-