ENDOLOGIX AFX; BIFURCATED STENT GRAFT Back to Search Results
Model Number BA25-110/I16-30
Medical Device Problem Codes Material Puncture/Hole (1504); Product Quality Problem (1506); Material Integrity Problem (2978)
Health Effect - Clinical Codes Death (1802); Failure of Implant (1924); Rupture (2208)
Date of Event 04/16/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.If additional information is obtained at a later time that is pertinent to this event, a follow-up report will then be submitted.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the afx abdominal aortic aneurysm stent.Approximately three (3) years post initial procedure, the patient presented emergently with a ruptured aaa.The ct scan revealed a type iiia endoleak and a potential iiib endoleak.The patient was transferred to another hospital for treatment but expired in transit.
 
Additional Manufacturer Narrative
A clinical evaluation of the reported event could not be completed due to a lack of receipt of relevant medical records and or imaging; requests were made and denied for medical records, and no response was received for medical imaging.As such, event determination, off-label conditions, related patient harms, and patient disposition could not be independently assessed.The manufacturing lot review confirmed all devices met specifications prior to release.Endologix continues to investigate this event and similar events to ensure the highest quality and patient safety.Device iteration is afx with duraply.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key8584703
Report Number2031527-2019-00222
Device Sequence Number14210617
Product Code MIH
UDI-Device Identifier00818009012180
UDI-Public(01)00818009012180(17)19111
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/11/2019
Device Model NumberBA25-110/I16-30
Device Lot Number1414301-001
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/23/2019
Supplement Date Received by Manufacturer04/23/2019
Initial Report FDA Received Date05/06/2019
Supplement Report FDA Received Date07/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0006-2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX-LIMB STENT GRAFT:1253156-020; AFX-LIMB STENT GRAFT:1350053-006; AFX-VELA SUPRARENAL:1286549-038
Outcome Attributed to Adverse Event Death;
Patient Age58 YR
-
-