| Model Number |
2002 |
| Medical Device Problem Code |
Failure to Charge (1085)
|
| Health Effect - Clinical Code |
Therapeutic Response, Decreased (2271)
|
| Date of Event |
04/04/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient visited the clinic on (b)(6) 2019 complaining of a flashing light on the mobile charger with an error code 25 displayed.The clinic reached out to reshape lifesciences field clinical specialist to report the issue and the field clinical specialist asked to have the patient brought back in for further investigation.On (b)(6) 2019, the patient was seen by the clinic.The device lead impedance on the anterior tip to posterior tip was 1583 ohms.With an impedance above 1500 ohms, therapy will not be delivered.The clinic suggested the patient have a complete removal of the device.The removal of the device is scheduled for (b)(6) 2019.
|
| |
|
Event or Problem Description
|
|
The patient visited the clinic on (b)(6) 2019 complaining of a flashing light on the mobile charger with an error code 25 displayed.The clinic reached out to reshape lifesciences field clinical specialist to report the issue and the field clinical specialist asked to have the patient brought back in for further investigation.On (b)(6) 2019, the patient was seen by the clinic.The device lead impedance on the anterior tip to posterior tip was 1583 ohms.With an impedance above 1500 ohms, therapy will not be delivered.The clinic suggested the patient have a complete removal of the device.The removal of the device is scheduled for on (b)(6) 2019.
|
| |
|
Search Alerts/Recalls
|