| Catalog Number |
7210812F |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a hip arthroscopy the generator was not working.No back up device was available, no delay nor patient injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of functional failure could not be reproduced.Unit passed functional testing during 6 hour burn in and power output was normal throughout testing.All functions perform as expected.All power and impedance readings were within spec.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
| |
|