MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383714
Device Problems Leak/Splash (1354); Obstruction of Flow (2423)
Patient Problem No Information (3190)
Event Date 04/10/2019
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that during use of the bd pegasus¿ safety closed iv catheter system the flow was occluded.Foreign complaint the following information was provided by the initial reporter, translated from (b)(6) to english: event date: (b)(6) 2019.The nurse found that the extension tubing leaked at the luer connection during flushing after penetration.The catheter was removed and replaced.The new catheter worked normally.
 
Manufacturer Narrative
Investigation: a device history review was conducted for lot number 8256300.Our records show that this is the 6th instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.Only a photo was provided for the investigation for this failure mode, based off the photo and history investigation information for this related defect, the possible root cause is process related.The flaring process is when the metal wedge crashes the support platform/ supporting gripper and causes the metal wedge to have damage.The damaged metal wedge scratches the extension tubing and causes the leaking issue.The manufacturing site has taken corrective actions to prevent this failure mode from occurring such as checking the alignment and cam angle in weekly maintenance occurrences and making machine repairs once any damage is found.
 
Event Description
It was reported that during use of the bd pegasus¿ safety closed iv catheter system the flow was occluded.Foreign complaint the following information was provided by the initial reporter, translated from chinese to english: event date: (b)(6)2019.The nurse found that the extension tubing leaked at the luer connection during flushing after penetration.The catheter was removed and replaced.The new catheter worked normally.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
MDR Report Key8586979
MDR Text Key145114623
Report Number8041187-2019-00351
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
PMA/PMN Number
PENDING
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 06/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Catalogue Number383714
Device Lot Number8256300
Initial Date Manufacturer Received 04/26/2019
Initial Date FDA Received05/07/2019
Supplement Dates Manufacturer Received04/26/2019
Supplement Dates FDA Received06/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-