| Model Number |
PBX-SYS-AFFIRM-3D |
| Medical Device Problem Code |
Imprecision (1307)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As of today, the investigation is still in process and a follow up will be filed as needed.
|
| |
|
Event or Problem Description
|
|
It was reported that "needle hit array and now causing artifact." additional information obtained noted that during a breast biopsy when the needle fired an odd sound was heard and the patient had some discomfort.The patient was numbed more and the biopsy was completed successfully.No medical intervention was needed.A field engineer was dispatched to the site.
|
| |
|
Event or Problem Description
|
|
It was determined that the needle stroke margin needed to be corrected.Once this was completed by an applications specialist the system was working as intended.
|
| |
|