| Model Number |
67PFPS45 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/09/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Possible lot numbers (d4): 3700959, 3708636.
|
| |
|
Event or Problem Description
|
|
Information was received that a smiths medical portex bivona flextend tts pediatric tracheostomy tube was placed in a patient after a routine tracheostomy (trach) tube change.When the tube was placed, it was reported that "the sheath over the obturator moved up and jammed so that the obturator was not able to be taken out".Subsequently, the trach had to be taken out and the old device was put back into the patient.There were no adverse patient effects.Information was received that a smiths medical portex bivona flextend tts pediatric tracheostomy tube was placed in a patient after a routine tracheostomy (trach) tube change.When the tube was placed, it was reported that "the sheath over the obturator moved up and jammed so that the obturator was not able to be taken out".Subsequently, the trach had to be taken out and the old device was put back into the patient.There were no adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation results: two photographs of bivona tracheostomy tubes were received for investigation.No damage could be detected in the obturators.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.No root cause could be determinated since the complaint was not confirmed.
|
| |
|
Search Alerts/Recalls
|