| Catalog Number |
71654501 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.A small section of the device fractured off at the guide connection point and was not returned.The device was manufactured in 2016 and exhibits signs of extensive wear/ usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Credit will be issued for the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the product broke during impaction.No harm to patient or staff.Nothing got in patient.No procedural delays because smith & nephew back up was available.Patient survived.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.A small section of the device fractured off at the guide connection point and was not returned.The device was manufactured in 2016 and exhibits signs of extensive wear/ usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|
| |
|
Search Alerts/Recalls
|