• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LEGION HINGE FEM TRL SZ 5 LT; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LEGION HINGE FEM TRL SZ 5 LT; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY Back to Search Results
Catalog Number 71434173
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/11/2019
Event Type  malfunction  
Event Description
It was reported that during procedure the 2mm offset coupler is stuck in the 5l hinge femoral trial.No injuries or delays reported.Back up available.
 
Manufacturer Narrative
The associated devices were returned and evaluated.A visual inspection of the returned devices confirmed the stated failure mode.The lock screw was missing from the coupler trial, which would not allow it to be removed from the femoral trial.The devices were manufactured in 2008 and 2012 and exhibit signs of extensive wear/usage.A review of complaint history did not reveal additional complaints for the listed batches for the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION HINGE FEM TRL SZ 5 LT
Type of Device
PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8592015
MDR Text Key144453591
Report Number1020279-2019-01816
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00885556032664
UDI-Public00885556032664
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 05/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71434173
Device Lot Number12MPG0001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2019
Date Manufacturer Received04/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-