|
The associated flexible shaft and the tray lids were returned and evaluated.Visual inspection of the returned devices showed that the flexible reamer shaft fractured within the shaft via mechanical overload.If sufficient offset bending or torsional forces are applied to the reamer shaft during use that exceed the strength of the material, a mechanical overload fracture can occur.The tray lids are bent, confirming the stated failure.The lids were manufactured in 2014 and the reamer shaft was manufactured in 2017.These instruments are reusable devices that can be exposed to numerous surgeries; damage from repeated use can occur.Several potential factors that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to/after each use and cleaning.Documentation review not performed for this event, as no manufacturing, packaging or labelling defects were associated with the reported failure.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|