| Catalog Number |
J207G |
| Medical Device Problem Code |
Component Incompatible (1108)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Additional evaluation summary: an unopened sample of product code j207, lot mk6373 was returned for analysis.During the visual inspection of the unopened sample, no defects were found on the package.The sample was opened and the spool had two holes punched and for product code j207 should be only one hole punched.Due to mismatch at the spool a characterization of the suture was performed and meet the requirements for a vicryl suture diameter 0.According to the sample condition, the assignable cause of the mix reel is due a manufacturing defect.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that the internal packaging of suture does not match the box packaging or packaging of suture.There is size 0 on the box, size 0 on the suture packaging, but suture is 2-0.The procedure was completed successfully.No adverse patient consequences reported.
|
| |
|