Manufacturing review: manufacturing records were reviewed (dhr, mrr¿s, scrap and mfg.Process changes) and there were not found evidence that the failure mode reported in this complaint is caused by the manufacturing process.Investigation summary: one 14.5 fr d/l glidepath 19 cm str hemodialysis catheter with surecuff kit was returned for evaluation.Visual and microscopic evaluations were performed.The investigation is confirmed for damaged/defective tunneler, as a break was noted on connecter of tunneler.The exact root cause for the damaged tunneler could not be determined.However, it is likely that supplier issues may have contributed to reported event.Labeling review: as this complaint has been found to be manufacturing/supplier related, a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) is not applicable.Product labeling would not have prevented this issue.Expiry date: 07/2020.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|