| Model Number |
27050E |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
03/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Customer stated the electrode was a cook (b)(4) cutting loop and was used with a karl storz working element.The working element was evaluated by our field service technician, and confirmed by customer, that it most likely had been repaired by a 3rd party, to the material, shaft, collar and white block.
|
| |
|
Event or Problem Description
|
|
Allegedly, during a transurethral removal of a prostate tumor, when the doctor seeded the loop into the white block to lock the electrode, it popped out.He tried a 2nd time and it popped out again, on the 3rd try when it popped out, it pulled and dislodged the urethra.An internal catheter was placed in the patient.
|
| |
|