• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. TRIVEX SYSTEM RESECTOR HANDPIECE; VARICOSE VEIN ALBATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. TRIVEX SYSTEM RESECTOR HANDPIECE; VARICOSE VEIN ALBATION SYSTEM Back to Search Results
Catalog Number 7210387F
Medical Device Problem Code No Device Output (1435)
Health Effect - Clinical Code No Information (3190)
Date of Event 04/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have received the complaint device for evaluation.When the handpiece was connected into a control unit, the indicator light on the control unit flashed orange indicating an issue with the handpiece.We were unable to rotate the drive shaft when either the run or the window lock button was pressed.Upon disassembly, we did not noted any component failure and the adhesive on the pc board appeared to be in properly working condition.The seal housing was observed to have heat discoloration.The likely root cause of this issue is a seized ball bearing.Over time, particulates and water could enter into the ball bearing assembly since it is not a hermetically sealed component.We have already implemented a corrective and preventive action (capa) to resolve this issue by replacing the ball bearings with slip bearings which better prevent the bearings from corrosion and from seizing.This device was manufactured prior to the implementation of these changes.There was no harm to the patient as the result of this incident.The procedure was completed using a different handpiece.
 
Event or Problem Description
The resector failed to rotate during pre-use check.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIVEX SYSTEM RESECTOR HANDPIECE
Common Device Name
VARICOSE VEIN ALBATION SYSTEM
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 02148
MDR Report Key8593242
Report Number1220948-2019-00054
Device Sequence Number2169957
Product Code DWQ
UDI-Device Identifier00840663106561
UDI-Public00840663106561
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K032387
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number7210387F
Device Lot NumberYQ01541
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2019
Initial Report FDA Received Date05/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-