DEPUY MITEK LLC US 4580 FMS DUO+ PUMP/SHAVER COMBO -NS; DISTENSION UNIT, FLUID, ARTHROSCOPIC
|
Back to Search Results |
|
Catalog Number 284580 |
Device Problem
Suction Failure (4039)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/18/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported by the sales rep via phone that during an unspecified surgical procedure, it was observed that the fms duo + pump had poor outflow suction.They were able to complete the case using this unit with no delay or patient harm.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The device was received at service center and evaluated.Operational and diagnostic analysis does not confirm the reported issue (poor outflow suction).The testing of the unit was completed.The unit passed all functional tests and is fully operational.Since there was no defect found the root cause for the reported failure cannot be determined.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 07-02-2018 and passed all functional testing before being returned to the customer.No further investigation is required.At this point in time no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
|
|