• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 4580 FMS DUO+ PUMP/SHAVER COMBO -NS; DISTENSION UNIT, FLUID, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 4580 FMS DUO+ PUMP/SHAVER COMBO -NS; DISTENSION UNIT, FLUID, ARTHROSCOPIC Back to Search Results
Catalog Number 284580
Device Problem Suction Failure (4039)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/18/2019
Event Type  malfunction  
Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly.Udi: (b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported by the sales rep via phone that during an unspecified surgical procedure, it was observed that the fms duo + pump had poor outflow suction.They were able to complete the case using this unit with no delay or patient harm.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The device was received at service center and evaluated.Operational and diagnostic analysis does not confirm the reported issue (poor outflow suction).The testing of the unit was completed.The unit passed all functional tests and is fully operational.Since there was no defect found the root cause for the reported failure cannot be determined.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 07-02-2018 and passed all functional testing before being returned to the customer.No further investigation is required.At this point in time no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4580 FMS DUO+ PUMP/SHAVER COMBO -NS
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8594855
MDR Text Key145115552
Report Number1221934-2019-57103
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020607
UDI-Public10886705020607
Combination Product (y/n)N
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 04/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number284580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2019
Date Manufacturer Received06/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-