• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MO016 MO-MOBERLY HOT PACK, LARGE, 6X9IN; PACK, HOT OR COLD, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MO016 MO-MOBERLY HOT PACK, LARGE, 6X9IN; PACK, HOT OR COLD, DISPOSABLE Back to Search Results
Model Number 11443-012
Health Effect - Clinical Code Injury (2348)
Date of Event 01/01/1980
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device history record reviews were completed on the reported lots v9c314 and v9c381.The lots were found to have been manufactured and released to predetermined specifications.No anomalies were found during review of the records.At the time of this investigation, no samples were provided to cardinal health.Therefore; without a sample we were not able to perform any type of investigation in order to determine a root cause for the issue reported.Based on analyses conducted, the site has taken several measures and is continuously monitoring the process to help identify failure modes and implement corrective actions.A quality alert has been issued to create more awareness regarding the failure mode.All production employees have been trained on this issue.The site has increased the visual sampling size to ensure defects are being caught prior to being packaged.The manufacturing site has modified the machine parameter sheets and re-trained operators to ensure the brush and duckbill dive position is correct, which slows down the bubble when falling into the pouch and helps prevent bubble punctures on the pouch.
 
Event or Problem Description
Based on the report we received by the customer, the hot pack reportedly burst during activation, and one employee was treated in the emergency department with a saline flush to eyes, face and neck.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOT PACK, LARGE, 6X9IN
Common Device Name
PACK, HOT OR COLD, DISPOSABLE
Manufacturer (Section D)
MO016 MO-MOBERLY
808 w highway 24
moberly MO 65270
MDR Report Key8595040
Report Number1423537-2019-00315
Device Sequence Number2192910
Product Code IMD
UDI-Device Identifier00888499004880
UDI-Public00888499004880
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number11443-012
Device Catalogue Number11443-012
Device Lot NumberV9C314; V9C381
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2019
Initial Report FDA Received Date05/09/2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-