| Catalog Number |
15-616 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The reason for reoperation is rupture.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side possible rupture.Additional events of "pain and hardness" and "droopy" stating that it may be due to the implanting physician "not removing enough tissue." device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: a4, b5, d6b, h6.
|
| |
|
Event or Problem Description
|
|
The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the product analysis performed, the assessments of the complaints are: ¿ rupture: observed three openings assessed as surgical damage.¿ pain: unable to observe as it is not related to the manufacturing process.¿ visibility/palpability: unable to observe as it is not related to the manufacturing process.No other observations observed, no further actions are required.Additional, changed, and/or corrected data: b5, d9, h3, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side possible rupture.Additional events of "pain and hardness" and "droopy" stating that it may be due to the implanting physician "not removing enough tissue." healthcare professional later reported "smaller implant on the patient's left." device has been explanted and replaced with non-abbvie device.
|
| |
|