• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 15 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 15 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 15-616
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The reason for reoperation is rupture.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
 
Event or Problem Description
Healthcare professional reported left side possible rupture.Additional events of "pain and hardness" and "droopy" stating that it may be due to the implanting physician "not removing enough tissue." device remains implanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: a4, b5, d6b, h6.
 
Event or Problem Description
The device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: based on the product analysis performed, the assessments of the complaints are: ¿ rupture: observed three openings assessed as surgical damage.¿ pain: unable to observe as it is not related to the manufacturing process.¿ visibility/palpability: unable to observe as it is not related to the manufacturing process.No other observations observed, no further actions are required.Additional, changed, and/or corrected data: b5, d9, h3, h6.
 
Event or Problem Description
Healthcare professional reported left side possible rupture.Additional events of "pain and hardness" and "droopy" stating that it may be due to the implanting physician "not removing enough tissue." healthcare professional later reported "smaller implant on the patient's left." device has been explanted and replaced with non-abbvie device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 15 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
suite 100
irvine, CA 92612
8479366324
MDR Report Key8595339
Report Number9617229-2019-04685
Device Sequence Number2293633
Product Code FTR
UDI-Device Identifier10888628001541
UDI-Public10888628001541
Combination Product (Y/N)N
Initial Reporter StateFL
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/11/2019
Device Catalogue Number15-616
Device Lot Number2670652
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2019
Supplement Date Received by Manufacturer01/03/2025
03/20/2025
Initial Report FDA Received Date05/09/2019
Supplement Report FDA Received Date01/22/2025
03/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/11/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight57 KG
-
-