| Catalog Number |
F2832 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a dental procedure on unknown date in (b)(6)2019 and suture was used.The suture broke during the passage in tissue before making the second knot.Another device from competitor was used to complete the procedure.There were no adverse patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: it was reported performance breakage suture.One empty labeled winding former and a needle-suture piece of product code f2830, lot mlh139 were returned for analysis.During visual inspection of the needle-suture piece, the swage and attachment area were noted to be as expected, and the end of suture piece was observed cut appears to be by use of a surgical instrument.Additionally, the functional test could not be performed.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Due to the sample condition, the assignable cause of performance breakage suture, suggests an improper handling of the sample.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A manufacturing record evaluation was performed for the finished device batch mlh139, xvf283019 and no non-conformances were identified.
|
| |
|