|
Product complaint # (b)(4).Investigation summary: it was reported: suture breakage.Two unopened samples of product code f2830, lot mlh139 was returned for analysis.During the visual inspection of two unopened samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the representative samples, no suture breakage was found and the tested samples met the finished goods requirements.One empty labeled winding former and a needle-suture piece of product code f2830, lot mlh139 were returned for analysis.During visual inspection of the needle-suture piece, the swage and attachment area were noted to be as expected, and the end of suture piece was observed cut, which appears to be by use of a surgical instrument.Additionally, the functional test could not be performed.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Due to the sample condition, the assignable cause of the suture breakage suggests an improper handling of the sample.
|