BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.75/18; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 364489
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/10/2019
Type of Reportable Event Malfunction
Event or Problem Description
Ous mdr - an orsiro drug-eluting stent system was selected to treat a calcified coronary lesion.It was reported that the stent dislodged and was detected at the level of the valve of the used introducer sheath.
 
Additional Manufacturer Narrative
The technical investigation revealed that the balloon is well folded and does not show any signs of inflation.Microscopic analysis of the balloon surface revealed clear visible stent imprints, indicating that the stent was properly crimped in between the two radiopaque markers at the time of delivery.The stent is severely deformed over its entire length.Review of the manufacturing history of the product detailed above did not reveal any non-conformity.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.The device in question was manufactured according to specifications and successfully passed all inprocess controls as well as the final inspection.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.75/18
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key8597374
Report Number1028232-2019-01919
Device Sequence Number2074532
Product Code NIQ
UDI-Device Identifier07640130439047
UDI-Public7640130439047
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/09/2020
Device Model Number364489
Device Catalogue NumberSEE MODEL NO.
Device Lot Number09184391
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2019
Initial Date Received by Manufacturer 05/09/2019
Supplement Date Received by Manufacturer07/17/2019
Initial Report FDA Received Date05/09/2019
Supplement Report FDA Received Date07/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-