| Model Number |
EXPRESS |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio determined that the cause of the reported issue was due to a shorted capacitor, designator c76, on the digital pcb assembly.The customer received a replacement device.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report a non-critical issue with their device.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that the device unexpectedly powered off while it was being tested.
|
| |
|