STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HARH45
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Codes Death (1802); Blood Loss (2597)
Date of Event 04/03/2019
Type of Reportable Event Death
Additional Manufacturer Narrative
The device was not returned to stryker sustainability solutions for evaluation.The facility stated they are working with their risk management team.The device is being held by the facility.As the device was not returned for evaluation, inspection was unable to be performed.A review of the dhr could not be performed as the lot number and/or serial number was not reported.Therefore, the most likely root causes are: user expectation of adaptive tissue technology functionality; user selects improper min settings on generator; ancillary equipment issues (e.G.Generator, handpiece, and/or adaptor); user misinterprets generator signal; tissue accumulation between the blade and shaft.The instructions for use (ifu) state: care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Clamping the tissue pad against the active blade without tissue on the full length of the blade will result in higher blade, clamp arm and distal shaft temperatures and can result in possible damage to the instrument.If this occurs, there may be an instrument failure, and the generator touchscreen displays a troubleshooting message.Incidental and prolonged activation against solid surfaces, such as bone, may result in blade heating and subsequent blade failure, and should be avoided.If activation is unintentionally stopped while sealing, maintain jaw closure and reactivate.The instrument can be used for dissection, grasping, coagulation, and cutting between the blade and clamp arm.Note: to achieve complete sealing, the trigger should be fully closed and the vessel fully contained between clamp arm and blade of device.An audible and tactile "click" indicates full trigger closure.To achieve full closure of the jaws of the device, squeeze the plastic trigger until you feel it stop against the plastic handle (plastic to plastic).If full trigger closure is released prior to or during activation on tissue, an audible and tactile "click" is evident.Increase grip force until full trigger closure is achieved.For optimal performance and to avoid tissue sticking, clean the instrument blade, clamp arm, and distal end of the shaft throughout the procedure by activating the instrument tip in saline.Note: do not touch the instrument to metal while activated.Note: do not clean the blade tip with abrasives.It can be wiped with a moist gauze sponge to remove tissue, if necessary.If tissue is still visible in the clamp arm, use hemostats to remove residue, taking care not to actuate the hand piece.If desired, the instrument may be unplugged.The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the advanced hemostasis hand control button.Do not attempt to seal vessels in excess of 7 mm in diameter.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
 
Event or Problem Description
It was reported the harmonic scalpel (harh45) failed, it was not sealing properly.This was noted post operation.It was stated the vessel was fully sealed during the bariatric procedure, the patient was released, but came back and the vessel was bleeding and the patient expired.These are commonly used devices that are readily available.
 
Event or Problem Description
It was reported the harmonic scalpel (harh45) failed, it was not sealing properly.This was noted post operation.It was stated the vessel was fully sealed during the bariatric procedure, the patient was released, but came back and the vessel was bleeding and the patient expired.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
On 11/18/2020, stryker sustainability solutions became aware that mdr section b2.(outcomes attributed to ae) was not completed for this mdr.This supplemental mdr serves to provide this information.The reported event will continue to be monitored through post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
MDR Report Key8597918
Report Number0001056128-2019-00026
Device Sequence Number14001307
Product Code NLQ
UDI-Device Identifier07613327346244
UDI-Public07613327346244
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K161693
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberHARH45
Device Catalogue NumberHARH45RR
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/12/2019
Supplement Date Received by Manufacturer11/18/2020
Initial Report FDA Received Date05/09/2019
Supplement Report FDA Received Date12/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
-
-