| Model Number |
MMT-523LNAB |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer's mother reported via phone call that the insulin pump had a motor position encoder error.The customer's blood glucose value was 185 mg/dl at the time of the incident.The customer reported that while changing the set, the insulin pump stopped working and had a motor error alarm.The customer reported that the drive support cap appeared normal or slightly indented.The insulin pump was not exposed to high magnetic field.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the displacement test, rewind, basic occlusion test, occlusion test, prime test and excessive no delivery test or motor position encoder error alarm due to motor error alarms.
|
| |
|