| Catalog Number |
03P36-25 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complete patient identification: (b)(6).An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported a false elevated afp initial result when processing on the architect i2000sr.The initial result was 581.87 and retest was 3.28 ng/ml.A new blood draw generated a result of 3.2 ng/ml.A previous afp result from 3 months prior was 3 ng/ml.Additional diagnostic testing was provided for this patient: (b)(6).Alt was 70 and ast was 40.No impact to patient management was reported.
|
| |
|
Event or Problem Description
|
|
Addtional information from the customer stated that on admission the patient underwent an mri examination and the patient was discharged without any particular medication or treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
Event was updated to include additional information received from the customer regarding the event.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed for the customer issue and included a review of the complaint text, testing of a retained kit of lot 91429fn00, a search for similar complaints, a review of trending data, a review of the historical performance for reagent lot 91429fn00, a review of product quality history, and a review of product labeling.Testing was performed using an in-house retained kit of lot 91429fn00 and all specifications were met indicating the lot is preforming acceptably.The ticket searches determined normal complaint activity for the likely cause lot.Complaint trending report review determined that there is no non-statistical or adverse trend for the product for the complaint issue.World wide data was used to review the historical performance of lot 91429fn00 and determined the patient median result for the lot is comparable with all other lots in the field and confirms no systemic issue for the lot.A review of the product quality history for the lot did not identify issues associated with the customer observation.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation a product deficiency was not identified.
|
| |
|
Search Alerts/Recalls
|