Age at time of event: 18 year or older.Device evaluated by manufacturer: the device was returned for analysis.The distal shaft was clotted with blood.Blood was removed during the lab analysis.The tip, distal shaft, collar and hypotube were microscopically and visually inspected.Inspection revealed a kink in the hypotube located 6.1 cm from the strain relief, collar damage (apex of collar flared upwards), and tip damage (torn/misshapen).Inspection of the rest of the device found no other damage or irregularities.Functional testing was performed with a test stented balloon catheter.The stented device was loaded into the guidezilla collar, and advanced through the tip without issue.The collar of the device was measured was within specification.The tip could not be accurately measured due to the tip damage.
|
Reportable based on device analysis completed on (b)(6) 2019.It was reported that the stent dislodged inside the catheter.The target lesion was located in the moderately tortuous proximal left anterior descending artery.A guidezilla ii guide extension catheter was selected for use.During procedure, an unknown stent was unable to cross the lesion, thus it was deilvered in combination with the guidezilla catheter.The stent was tried to advance as it was into the patient's body even if there was resistance felt at the port part of the catheter.Subsequently,it was noted that the stent dislodged inside the collar part of the catheter under fluoroscopy.The dislodged stent was then removed together with the catheter.The procedure was completed with this device.No patient complications were reported.However, analysis revealed that the apex of the collar flared upwards.
|