Catalog Number 254401006 |
Device Problems
Break (1069); Crack (1135); Use of Device Problem (1670)
|
Patient Problem
Not Applicable (3189)
|
Event Date 04/22/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
During a total knee replacement the femoral impactor cracked when used to impact the femoral trial into position.A small fragment was missing from the main body of the instrument and was unable to be found.The instrument was not inspected for cracks immediately before use and the surgical site was thoroughly examined and determined to be free of plastic debris before the case continued.No ae to patient.
|
|
Manufacturer Narrative
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: examination of the returned device confirms the complaint; the impactor has cracked and had a piece missing.Root cause attributed to product design.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
|