Model Number 3660
Device Problem Difficult to Insert (1316)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/19/2019
Event Type  malfunction  
Manufacturer Narrative
Analysis confirmed the issue.Microscopic inspection of the header chamber for port 1 thru 8 identified an obstruction.The device history record was reviewed and found to be complete.The obstruction is excess silicone that is preventing the lead from being fully inserted.
 
Event Description
It was reported a lead was unable to be inserted into the header of the ipg during a procedure to implant an scs system.Another ipg was used to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8601240
MDR Text Key144745496
Report Number3006705815-2019-01760
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020192
UDI-Public05415067020192
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/11/2020
Device Model Number3660
Device Lot NumberA000070123
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/12/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-