Catalog Number C01B
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Numbness (2415)
Date of Event 04/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
X-ray image review results: single lateral image of t6 kyphoplasty shows extravasation of cement from the posterior aspect of the vertebral body.By report, it is midline into the spinal canal.There is a substantial height loss of the vertebrae.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent kyphoplasty at t6 due to vertebral compression fracture.Intra-op, during cement injection into the vertebral body, cement was noticed to be leaking from the posterior side into spinal canal.When the issue was noticed, the physician stopped the use of cement immediately.Post-op, the patient experienced numbness in feet as a result of this event.Cement was mixed for about 3 minutes and it was doughy and homogeneous prior to delivery into the patient.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON HV-R BONE CEMENT
Common Device Name
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key8602220
Report Number1030489-2019-00525
Device Sequence Number13998414
Product Code NDN
UDI-Device Identifier00643169097834
UDI-Public00643169097834
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K180700
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Catalogue NumberC01B
Device Lot Number0009517077
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/11/2019
Supplement Date Received by Manufacturer04/11/2019
Initial Report FDA Received Date05/11/2019
Supplement Report FDA Received Date05/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age67 YR
-
-