• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DII CONTROLLER; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DII CONTROLLER; ARTHROSCOPE Back to Search Results
Catalog Number 72200873
Device Problem Circuit Failure (1089)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
It was reported that a short circuit was detected.No case involved.
 
Manufacturer Narrative
A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with alarm sounding and "short circuit detected" error message on port a only.Cause of alarm and error is shorted electronic components on the main pcb.Reason for electronic component failure is unknown but a shorted out mdu is a possibility.Product passed functional testing with a known good main pcb installed.The complaint investigation has concluded the cause of the failure to be defective electronic components on the main digital pcb.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DII CONTROLLER
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key8603748
MDR Text Key144960756
Report Number1643264-2019-00347
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200873
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2019
Initial Date FDA Received05/13/2019
Supplement Dates Manufacturer Received06/13/2019
Supplement Dates FDA Received06/14/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-