Catalog Number 74027247
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/17/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to suspect superficial infection.Debridement and insert replace were performed.No more information provided yet.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A clinical evaluation was conducted and no relevant clinical medical information was provided to conduct a thorough medical assessment.Should any additional clinical information be provided this complaint will be re-assessed.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This included a review of sterilization documents which indicated that the product was sterilized according to sterilization release documentation.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS XLPE ART ISRT SZ 3-4 LT 18MM
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8603858
Report Number1020279-2019-01886
Device Sequence Number16953031
Product Code JWH
UDI-Device Identifier00885556177914
UDI-Public00885556177914
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number74027247
Device Lot Number18EM04056
Initial Date Received by Manufacturer 04/19/2019
Supplement Date Received by Manufacturer04/19/2019
Initial Report FDA Received Date05/13/2019
Supplement Report FDA Received Date10/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PART NUMBER 71932635 / LOT NUMBER 18MM13457; PART NUMBER 74022123 / LOT NUMBER 18HM09450; PART NUMBER 74022223 / LOT NUMBER 18HM23555
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age70 YR
-
-