• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LEGION PS OXIN FEM SZ7 RT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LEGION PS OXIN FEM SZ7 RT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Catalog Number 71421207
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 04/19/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to severe metallosis.
 
Additional Manufacturer Narrative
The associated complaint devices were returned and evaluated.A lab analysis was conducted and the returned devices were examined visually.No destructive analysis was performed.The returned components are damage, has deformation, gouging and scratches were observed on the articular surface of the femoral component.The articulating surface of the tibial insert was also damaged and scratched.These features observed on the tibial insert have been reported as being created by third-body particulate (bone cement, bone, etc.) in the joint space.Cement and bone particulate are known to be produced during a total knee arthroplasty (tka), even under carefully controlled conditions.Damage and scratches were observed on the tibial baseplate.Cement was observed on the tibial baseplate, femoral, and patella components.Scratches were also observed on the articulating surface of the patella component.No material or manufacturing deviations were observed during the investigation.A clinical evaluation was conducted and without the request clinical information a thorough medical investigation cannot be rendered.Should any additional clinical information be provided this complaint will be re-assessed.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This included a review of sterilization documents which indicated that the product was sterilized according to sterilization release documentation.Should information become available this complaint can be re-assessed.Available this complaint can be re-assessed.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION PS OXIN FEM SZ7 RT
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8604090
Report Number1020279-2019-01891
Device Sequence Number14040883
Product Code JWH
UDI-Device Identifier00885556029404
UDI-Public00885556029404
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K043440
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/11/2023
Device Catalogue Number71421207
Device Lot Number13LM09492
Initial Date Received by Manufacturer 04/19/2019
Supplement Date Received by Manufacturer12/10/2019
Initial Report FDA Received Date05/13/2019
Supplement Report FDA Received Date12/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age50 YR
Patient Weight85
-
-