STRYKER INSTRUMENTS-KALAMAZOO SILVERGLIDE PROSERIES BIPOLAR FORCEPS, BAYONET, 22CM / 1.5MM TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 6730220015
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/16/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during testing at the manufacturer facility insulation was missing near the tips of the device, posing the risk of a small component being lost in a surgical site.There were no adverse consequences related to this event.
 
Event or Problem Description
It was reported that during testing at the manufacturer facility insulation was missing near the tips of the device, posing the risk of a small component being lost in a surgical site.There were no adverse consequences related to this event.
 
Additional Manufacturer Narrative
Additional information: manufacture date follow-up report submitted to document device evaluation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILVERGLIDE PROSERIES BIPOLAR FORCEPS, BAYONET, 22CM / 1.5MM TIP
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key8604595
Report Number0001811755-2019-01660
Device Sequence Number14208313
Product Code GEI
UDI-Device Identifier04546540695963
UDI-Public04546540695963
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K992931
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number6730220015
Device Lot Number75817
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2019
Initial Date Received by Manufacturer 04/16/2019
Supplement Date Received by Manufacturer06/13/2019
Initial Report FDA Received Date05/13/2019
Supplement Report FDA Received Date06/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-