• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 RECLAIM PROX TRIAL SHAFT 95; RECLAIM INSTRUMENTS : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 RECLAIM PROX TRIAL SHAFT 95; RECLAIM INSTRUMENTS : FEMORAL TRIALS Back to Search Results
Catalog Number 297529095
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during reclaim hip revision, several instruments height segments were damaged and neck trial screws stripped.A screw in the dome of a 40mm liner also dislodged.No surgical delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned.A provided photograph confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM PROX TRIAL SHAFT 95
Common Device Name
RECLAIM INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key8605556
Report Number1818910-2019-93347
Device Sequence Number3236561
Product Code LXH
UDI-Device Identifier10603295156284
UDI-Public10603295156284
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number297529095
Device Lot NumberNB12275
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/24/2019
Supplement Date Received by Manufacturer06/20/2019
Initial Report FDA Received Date05/13/2019
Supplement Report FDA Received Date06/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-