| Model Number |
NL5000 |
| Medical Device Problem Code |
Human-Device Interface Problem (2949)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
|
| Date of Event |
04/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The omnitom ct system has significant weight and size.We are releasing a service bulletin for customer that highlights people to get familiarized with the omnitom drive system and to refrain from maneuvering/driving the system in space restricted locations.
|
| |
|
Event or Problem Description
|
|
During navigation of the omnitom ct system, technologist was unable to get the omnitom to turn or do anything other than go backward.This resulted in the technologist getting backed up against a sink.After several minutes the technologist was able to get the system to turn however the ct system kept stopping while transporting it to the storage location.No patients were harmed a result of this incident however the technologist had some bruising.
|
| |
|