| Model Number |
39200 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: visual examination of the returned device revealed multiple kinks to the sheath and the inner liner.The distal end of the middle sheath was no longer in contact with the proximal end of the tip.The lock was on the rack, and the rack measured 6.3cm from the handle to the pull grip.The stent was sticking out approximately 1mm from the distal end of the middle sheath.Inspection of the remainder of the device presented no other damages or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that inadvertent deployment occurred.A 12x40x75mm epic stent was selected however, part of the stent was exposed when it was removed from the packaging.When attempted to fix, the stent was inadvertently deployed.It was not used in patient.
|
| |
|