U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC Back to Search Results
Model Number 39200
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/23/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: visual examination of the returned device revealed multiple kinks to the sheath and the inner liner.The distal end of the middle sheath was no longer in contact with the proximal end of the tip.The lock was on the rack, and the rack measured 6.3cm from the handle to the pull grip.The stent was sticking out approximately 1mm from the distal end of the middle sheath.Inspection of the remainder of the device presented no other damages or irregularities.
 
Event or Problem Description
It was reported that inadvertent deployment occurred.A 12x40x75mm epic stent was selected however, part of the stent was exposed when it was removed from the packaging.When attempted to fix, the stent was inadvertently deployed.It was not used in patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8605839
Report Number2134265-2019-05111
Device Sequence Number2510863
Product Code NIO
UDI-Device Identifier08714729805151
UDI-Public08714729805151
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P110035
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/27/2023
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0022718299
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2019
Initial Report FDA Received Date05/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-