| Catalog Number |
SDH29A |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/28/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.
|
| |
|
Event or Problem Description
|
|
It was reported that during a left hemicolectomy surgery, after fired, noted some staples malformed with irregular shape.Used suture to oversew.There was no patient consequence reported.No additional information can be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch r59m27.Investigation summary: the analysis results found that the sdh29a device arrived with no apparent damage.The breakaway washer was present and cut and there were no staples present, indicating that the device achieved a full firing stroke.In addition, the washer was noted to have nicks.The device was reloaded with staples, a new washer was placed on the device and it was tested for functionality.It fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.The event described could not be confirmed as the device performed without any difficulties noted.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|