• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ILS 29MM, STRAIGHT; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ILS 29MM, STRAIGHT; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number SDH29A
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # unk.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.
 
Event or Problem Description
It was reported that during a left hemicolectomy surgery, after fired, noted some staples malformed with irregular shape.Used suture to oversew.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Batch r59m27.Investigation summary: the analysis results found that the sdh29a device arrived with no apparent damage.The breakaway washer was present and cut and there were no staples present, indicating that the device achieved a full firing stroke.In addition, the washer was noted to have nicks.The device was reloaded with staples, a new washer was placed on the device and it was tested for functionality.It fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.The event described could not be confirmed as the device performed without any difficulties noted.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILS 29MM, STRAIGHT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key8605868
Report Number3005075853-2019-19060
Device Sequence Number1756707
Product Code GDW
UDI-Device Identifier10705036003533
UDI-Public10705036003533
Combination Product (Y/N)N
PMA/510(K) Number
K983536
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/10/2023
Device Catalogue NumberSDH29A
Device Lot NumberR40U3Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2019
Initial Date Received by Manufacturer 04/29/2019
Supplement Date Received by Manufacturer05/20/2019
Initial Report FDA Received Date05/13/2019
Supplement Report FDA Received Date06/14/2019
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1277-2019
Patient Sequence Number1
-
-