• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MATERIALISE NV TRUMATCH; MIDFACE/MANDIBLE - TITANIUM 3D PRINTED PLATE FOR MANDIBLE, MINI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MATERIALISE NV TRUMATCH; MIDFACE/MANDIBLE - TITANIUM 3D PRINTED PLATE FOR MANDIBLE, MINI Back to Search Results
Model Number SD980.016
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/11/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device was designed and manufactured according to specifications.There was no change in the cleaning or sterilisation process.Additional investigation is still ongoing.
 
Event or Problem Description
The patient underwent a bi-sagittal split osteotomy of the mandible.Post-operative an infection was reported on one side.
 
Additional Manufacturer Narrative
After concluding the investigation, a most likely root cause was identified concerning the proximity of the planned screws of the plates to the superior border of the mandible.
 
Event or Problem Description
The patient underwent a bi-sagittal split osteotomy of the mandible.Post-operative an infection was reported on one side.
 
Additional Manufacturer Narrative
Device was designed and manufactured according to specifications.There was no change in the cleaning or sterilisation process.Different design parameters have been investigated and are not considered as a possible root cause at this time.Manufacturing parameters have been investigated and are not considered as a possible root cause at this time.Additional investigation is still ongoing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUMATCH
Common Device Name
MIDFACE/MANDIBLE - TITANIUM 3D PRINTED PLATE FOR MANDIBLE, MINI
Manufacturer (Section D)
MATERIALISE NV
technologielaan 15
leuven, 3001
BE  3001
MDR Report Key8607130
Report Number3003998208-2019-00022
Device Sequence Number19403387
Product Code JEY
UDI-Device Identifier05420060380174
UDI-Public05420060380174
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K170272
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSD980.016
Device Catalogue NumberSD980.016
Device Lot NumberME18PIXROT
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/13/2019
Supplement Date Received by Manufacturer05/13/2019
11/27/2019
Initial Report FDA Received Date05/14/2019
Supplement Report FDA Received Date06/17/2019
12/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-