| Model Number |
SD980.016 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
02/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device was designed and manufactured according to specifications.There was no change in the cleaning or sterilisation process.Additional investigation is still ongoing.
|
| |
|
Event or Problem Description
|
|
The patient underwent a bi-sagittal split osteotomy of the mandible.Post-operative an infection was reported on one side.
|
| |
|
Additional Manufacturer Narrative
|
|
After concluding the investigation, a most likely root cause was identified concerning the proximity of the planned screws of the plates to the superior border of the mandible.
|
| |
|
Event or Problem Description
|
|
The patient underwent a bi-sagittal split osteotomy of the mandible.Post-operative an infection was reported on one side.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was designed and manufactured according to specifications.There was no change in the cleaning or sterilisation process.Different design parameters have been investigated and are not considered as a possible root cause at this time.Manufacturing parameters have been investigated and are not considered as a possible root cause at this time.Additional investigation is still ongoing.
|
| |
|