MAUDE Adverse Event Report: MEDTRONIC IRELAND SELECTSITE; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND SELECTSITE; CATHETER, PERCUTANEOUS Back to Search Results
Model Number C304S5906
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/23/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the implant procedure, the catheter's hub and first portion broke during slitting.The procedure completed.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the delivery catheter was returned and analyzed.The mechanical operation of the catheter slitting showed a spiral slit.The mechanical operation of the catheter shaft was broken in two pieces.The mechanical operation of the catheter shaft was kinked/buckled.Visual analysis of the catheter indicated damage during use.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELECTSITE
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway MN
MDR Report Key8607267
Report Number2182208-2019-00889
Device Sequence Number2092206
Product Code DQY
UDI-Device Identifier00613994729781
UDI-Public00613994729781
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K033989
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/05/2019
Device Model NumberC304S5906
Device Catalogue NumberC304S5906
Device Lot Number0008820972
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2019
Initial Date Received by Manufacturer 04/23/2019
Supplement Date Received by Manufacturer05/23/2019
Initial Report FDA Received Date05/14/2019
Supplement Report FDA Received Date05/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
-
-