| Model Number |
MT22608-BLU |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a "hwbbt" error was displayed on the receiver.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The receiver was returned for evaluation.An external visual inspection was performed and passed.A receiver charge and boot tests were performed and passed.The receiver data log was downloaded for review and hardware errors along with ¿screen error alarm¿ were observed.The receiver case was opened for an internal visual inspection and it passed.The reported event of an error icon display was confirmed during log review.A probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Search Alerts/Recalls
|