| Catalog Number |
06-2210 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Injury (2348); Joint Dislocation (2374)
|
| Date of Event |
04/16/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right hip was revised due to dislocation.Comparing implant sheets, a 22 +10 c-taper lfit head was revised to a 28 +10 c-taper lfit head.
|
| |
|
Search Alerts/Recalls
|