• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL INTRAFX ADVBR SCW10X30W/LGSHTH; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL INTRAFX ADVBR SCW10X30W/LGSHTH; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 254809
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/02/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint : (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore, unavailable for a physical evaluation.This complaint cannot be confirmed.No non-conformances were identified for this part number, lot number combination per qlik query executed on (b)(6) 2019.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi:(b)(4).The expiration date is unknown.
 
Event or Problem Description
It was reported by the sales rep via phone that during an acl reconstruction procedure the sales rep's 10x30 intrafix advance screw and 9x30 intrafix advance screw sheaths broke during insertion.The sales rep stated that the implants were fully removed from the patient with a hemostat.The implants were being used in conjunction with a size 9 graft, a 9.5 tunnel, and a large trial.The sales rep stated that the insertion of the implants was not off axis.The case was completed with another like-implant in the same bone hole with no patient harm or delay.The implants were discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAFX ADVBR SCW10X30W/LGSHTH
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key8609921
Report Number1221934-2019-57144
Device Sequence Number13976184
Product Code MAI
UDI-Device Identifier10886705026685
UDI-Public10886705026685
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K160804
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254809
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/02/2019
Initial Report FDA Received Date05/14/2019
Patient Sequence Number1
-
-