| Model Number |
1207-36 |
| Medical Device Problem Code |
Failure to Interrogate (1332)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that an asymptomatic patient presented at a follow-up in clinic.Upon interrogation, the implantable cardioverter defibrillator did not connect with radiofrequency (rf) telemetry.The rf antenna was unplugged and the device could be interrogated.The patient was in stable condition.
|
| |
|