| Catalog Number |
2C8607 |
| Medical Device Problem Code |
Failure to Disconnect (2541)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported there was a disconnection issue with an unspecified quantity of clearlink stopcock and i.V.Solution administration sets.The reporter stated that they could not remove the cap off the end.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information : the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection using the naked eye was performed in the photograph which noted that all components are correctly placed and according to product specifications.No defect was observed.By the nature of the sample, no additional tests were performed.The reported problem was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|