MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723LNAL
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/30/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Insulin pump received with intermittent button response due to flattened express bolus, esc, act, up and down button dome switch (creased).No button error alarm during testing.No unlocked j2/lcd keypad connector.Insulin pump passed self test.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had a keypad anomaly.The buttons were not responsive.The customer was on a ladder and the insulin pump fell.The customer was unable to prime the insulin pump.Customer's blood glucose level was not reported.Customer was advised to discontinue use of the device and revert to a backup plan.The customer was advised that the device would be replaced and agreed to not return the product for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8614507
Report Number3004209178-2019-97941
Device Sequence Number16899163
Product Code OYC
UDI-Device Identifier00643169478589
UDI-Public(01)00643169478589
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723LNAL
Device Catalogue NumberMMT-723LNAL
Device Lot NumberA4723LNALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2019
Initial Report FDA Received Date05/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age38 YR
Patient Weight98
-
-