| Model Number |
MMT-723LNAL |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/30/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Insulin pump received with intermittent button response due to flattened express bolus, esc, act, up and down button dome switch (creased).No button error alarm during testing.No unlocked j2/lcd keypad connector.Insulin pump passed self test.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had a keypad anomaly.The buttons were not responsive.The customer was on a ladder and the insulin pump fell.The customer was unable to prime the insulin pump.Customer's blood glucose level was not reported.Customer was advised to discontinue use of the device and revert to a backup plan.The customer was advised that the device would be replaced and agreed to not return the product for analysis.
|
| |
|